Clean operating room testing standards
Release time:2015-12-15 source:
Clean operating room testing standards
Clean operating room testing standards specification:
1, the cleansing operation should be located inside the fire resistance rating of not less than two buildings.
2, cleansing operation should be divided into separate fire district. When used with other parts in the same fire district, should take effective measures to separate fire and smoke, and should use refractory limit of not less than 2.00H separated from the other parts of the wall; and non-clean surgical region communicating door should use refractory limit of not less than B fire doors (except direct access to open outside the corridor outside the door or direct), or other appropriate measures for fire prevention technology.
3, technical dissection and operating rooms, auxiliary buildings and other clean room operation unit is connected through the site should take measures to fire and smoke, the fire resistance of its separate body of not less than 1.00H.
4. When the need to set the fire hydrant, fire hydrant may not be set up in the operating theater, but set fire hydrant outside the operating theater should be able to guarantee the two water cannons at the same time enrich the water column reaches the operating theater anywhere; when the clean room operation unit need when set fire hydrant , set fire hose reel and other fire-fighting facilities. When you need to set up automatic sprinkler system, may not be disposed in the operating room sprinklers. Cleanroom surgery unit should be set building fire extinguishers.
5, technical dissection should set fire clean room operation unit automatic alarm device.
6, clean room operation unit building fire safety regulations should be related to the building without windows windowless room within the building or smoke control system set up to design.
7, the exhaust port clean area should have anti-intrusion measures. Exhaust port must use the plate exhaust port.
8, exhaust valves clean area should be used for embedded installation, exhaust valve surface should be easy to clean and disinfect.
9, in a clean room surgery department should be able to set up the emergency cut off oxygen concentration dry pipe system.